The error message may appear under several conditions:
1. Incorrect balancing of the balance, check that the value of the unsheared TG signal is included in the selected measuring range.
2. internal scale problem which limits the measuring range (+/- 150 mg instead of +/- 200 mg for example) 3. Problem with the TG signal cord (not connected) or the TG signal acquisition module.
4. The current mass change is outside the measurement range.