With the primary vacuum option pushed: start with a purge (about 100 mbar to reach) and then go into primary vacuum pushed for 1h its value must be less than 5.10-2mbar. Stop the vacuum for 2 hours, the pressure must not be raised beyond 14mbar
KEP technologies Help Center is coming soon
With the primary vacuum option pushed: start with a purge (about 100 mbar to reach) and then go into primary vacuum pushed for 1h its value must be less than 5.10-2mbar. Stop the vacuum for 2 hours, the pressure must not be raised beyond 14mbar